Trio/Loop CGM Driver for Syai Ultra.
Currently Supported Models:
- Syai Ultra (X1), firmware V1.6 and V1.7
- Syai Ultra (X1), firmware V1.8 and V2.0 (supported, but not yet tested)
Note: In some countries an older sensor sold under the Syai Tag name (model number currently unknown) may still be available. This is not tested and may not be supported. Compatibility depends entirely on its firmware version. If you have one, please open a GitHub issue with the firmware/device version shown in the official Syai app so it can be looked into.
Syai sensors are read through your own Syai account, the same one you'd use in the official Syai app. You'll be asked to log in during setup. Trio talks to the sensor and your account directly. The official Syai app doesn't need to be involved once you're paired, and shouldn't be connected to the sensor at the same time as Trio (see Troubleshooting below).
During setup you'll also be asked whether to share the same usage data with Syai that their own app would normally send. Keeping this on helps if you ever need warranty or support from Syai; you can change this later in Settings → Account → Data Sharing.
In case an issue is not covered by the FAQ below please reach out on Trio Discord or open a GitHub issue.
Please make sure to attach your logs when you file a report. You can find them by going to the Syai settings menu and pressing the "Share Syai logs" button.
Q: There is a gap in my readings, what is happening?
A: The sensor's Bluetooth connection can drop occasionally; Trio reconnects automatically and backfills what it missed. If Settings shows a sensor error instead, the sensor itself has reported a fault and will need to be replaced.
Q: Can I use the official Syai app alongside Trio?
A: No. The sensor can only talk to one device at a time. It is not possible to run the Syai app along side Trio.
Q: Nightscout shows readings during warmup (often very high), or every reading twice. Is that a bug?
A: Not in SyaiKit. Trio never forwards warmup readings, so these come from Syai. If you linked Nightscout in the Syai Tag app, Syai's servers upload every reading Trio shares with them, including warmup. Use one Nightscout uploader, not both:
- Recommended: remove the Nightscout link in the Syai Tag app. Trio keeps uploading glucose along with everything else it sends, without warmup readings.
- Or turn off "Upload Glucose" in Trio's Nightscout settings. Syai becomes the only glucose uploader, but warmup readings will still appear.
For more information please join the Trio Discord: https://discord.triodocs.org/
SyaiKit is an independent, community-driven open-source project. It is not affiliated with, endorsed by, or supported by Syai Health Technology Pte. Ltd. The goal is interoperability with a device you already own, under your own account.
This is experimental software under active development. Automated insulin dosing is a safety-critical activity. Use at your own risk, keep a working fallback therapy available, and never dose on readings you have reason to doubt.
MIT — see LICENSE.